FDA Medical devices recall ·
Senographe Pristina recall
- Recall number
- Z-0327-2027
- Date
- (initiated August 28, 2026)
- Company / brand
- GE Medical Systems, LLC
3000 N Grandview Blvd, Waukesha, WI, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Initiated by
- FDA Mandated
Read the official FDA notice ↗
Product
Senographe Pristina
Why was it recalled?
GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
GTIN(s): 00840682118460, 00840682142052, 00840682145879, 00195278022745, 00195278276971, 00195278661999, 00195278665379, 00195278914026, 00198953109281, 10198953054311, 10195278901764.
Distribution
USA and OUS.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0327-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.