GE HealthCare OEC 9900 Elite C-arm Systems recall
- Recall number
- Z-2342-2025
- Date
- (initiated July 2, 2025)
- Company / brand
- GE Medical Systems, LLC
3000 N Grandview Blvd, Waukesha, WI, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Initiated by
- FDA Mandated
Read the official FDA notice ↗
Product
GE HealthCare OEC 9900 Elite C-arm Systems.
Why was it recalled?
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
GTIN: 00840682114349
Distribution
US Nationwide distribution.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.