FDA Medical devices recall ·

GE HealthCare OEC 9800 recall

Recall number
Z-2341-2025
Date
(initiated July 2, 2025)
Company / brand
GE Medical Systems, LLC
3000 N Grandview Blvd, Waukesha, WI, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Initiated by
FDA Mandated

Read the official FDA notice ↗

Product

GE HealthCare OEC 9800.

Why was it recalled?

GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

GTIN: 00840682114349

Distribution

US Nationwide distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2341-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.