FDA Medical devices recall ·

ENDOTAK RELIANCE G/SG with 4-SITE Connector… recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-2326-2025
Date
(initiated July 24, 2025)
Company / brand
Boston Scientific Corporation
4100 Hamline Ave N, Saint Paul, MN, United States
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
247,702 US; 344,670 OUS
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent Defibrillator Electrodes; Model Nos. 0282, 0283, 0285, 0286, 0292, 0293, 0295, 0296

Why was it recalled?

Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Model 0282 UDI-DI/GTIN 00802526431371; Model 0283 UDI-DI/GTIN 00802526431388 00802526504181; Model 0285 UDI-DI/GTIN 00802526431401; Model 0286 UDI-DI/GTIN 00802526431418; Model 0292 UDI-DI/GTIN 00802526431432 00802526474668 00802526480096 00802526485763 00802526504396 00802526531248 00802526541742 00802526552441 00802526611186; Model 0293 UDI-DI/GTIN 00802526431449 00802526531255 00802526541827 00802526552526 00802526554490 00802526611179; Model 0295 UDI-DI/GTIN 00802526431463 00802526474699 00802526531262 00802526541902 00802526554506 00802526611131; Model 0296 UDI-DI/GTIN 00802526431470 00802526474705 00802526480164 00802526480188 00802526531279 00802526541988 00802526552038

Distribution

Worldwide Distribution. US nationwide. International distribution worldwide.

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-2326-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.