FDA Medical devices recall ·

Nasopore Ex Firm 4cm fragmentable nasal dressing recall

Recall number
Z-1478-2025
Date
(initiated February 25, 2025)
Company / brand
Stryker Corporation
1941 Stryker Way, Portage, MI, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
182,344 total units
Hazard tags
Sterility
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Nasopore Ex Firm 4cm fragmentable nasal dressing, Catalog Number 5400-030-004 and Nasopore Forte plus 4cm Catalog Number 5400-030-004ITL

Why was it recalled?

There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

5400-030-004 UDI-DI: 07613154340835, Lot Numbers: 2024052333 2023012514 2023050514 2023080912 2024020214; 5400-030-004ITL UDI-DI: 07613327359442, Lot Numbers: 2023071933 2023102714 2024012314

Distribution

US Nationwide distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1478-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.