FDA Medical devices recall ·
Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry… recall
This is an FDA Class I recall — the most serious type.
- Recall number
- Z-1320-2026
- Date
- (initiated December 22, 2025)
- Company / brand
- Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States - Category
- Medical devices › Class I
- Classification
- Class I
- Status
- Ongoing
- Units affected
- 311
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81105RH 81172RH; Abbott Inquiry Steerable Diagnostic Catheter 20 Elec. 2-10-2mm 81202RH; Abbott Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81483RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81102RH 81104RH 81174RH 81531RH 81532RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 5mm 81107RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 2-5-2mm 81402RH 81404RH 81405RH 81474RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81473RH
Why was it recalled?
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
81105RH UDI-DI 10197344133253 Lot EP250618; 81172RH UDI-DI 10197344133284 Lot EP250618; 81202RH UDI-DI 10197344133307 Lot EP250618; 81483RH UDI-DI 10197344133444 Lot EP250618; 81102RH UDI-DI (case) 20197344133236¿ (ea) 10197344133239 Lot EP250618; 81104RH UDI-DI 10197344133246 Lot EP250618; 81174RH UDI-DI 10197344133291 Lot EP250618; 81531RH UDI-DI 10197344133512 Lot EP250618; 81532RH UDI-DI 10197344133529 Lot EP250618; 81107RH UDI-DI 10197344133260 Lot EP250618; 81402RH UDI-DI 10197344133352 Lot EP250618; 81404RH UDI-DI 10197344133376 Lot EP250618; 81405RH UDI-DI 10197344133383 Lot EP250618; 81474RH UDI-DI 10197344133437 Lot EP250618; 81473RH UDI-DI 10197344133420 Lot EP250618
Distribution
Distribution US nationwide.
What does Class I mean?
Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.
Source: FDA (medical devices), recall Z-1320-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.