FDA Medical devices recall ·
Medline medical procedure convenience kits labeled as: 1) LAVH PROCEDURE recall
- Recall number
- Z-1089-2026
- Date
- (initiated December 16, 2025)
- Company / brand
- Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 1457 units
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Medline medical procedure convenience kits labeled as: 1) LAVH PROCEDURE, REF CDS980754T; 2) GYN LAP HYSTERECTOMY CDS, REF CDS982662Q; 3) DAVINCI HYSTERECTOMY CDS, REF CDS985344F; 4) LAVH, REF DYNJQ9041R; 5) LAPAROSCOPIC HYSTERECTOMY, REF DYNJ900244I; 6) ROBOTIC LAVH, REF DYNJ901281I; 7) LAVH-LF, REF DYNJ901829T; 8) LEX LAP HYST, REF DYNJ902030J; 9) LITHOTOMY-SLINGS-LF, REF DYNJ905274D; 10) GYN LAPAROSCOPY, REF DYNJ905485J; 11) ROBOTIC HYSTERECTOMY, REF DYNJ906380C; 12) LAPAROSCOPIC HYSTERECTOMY, REF DYNJ910634; 13) LAVH GYN/ONC, REF DYNJ910927.
Why was it recalled?
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
REF CDS980754T: UDI/DI 10198459121814 (EA) 40198459121815 (CS), Lot Number 25EMD795; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25EBR237; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25EBP591; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25DBM353; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25CBL184; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25BBL633; REF CDS985344F: UDI/DI 10195327312565 (EA) 40195327312566 (CS), Lot Number 25FBK918; REF CDS985344F: UDI/DI 10195327312565 (EA) 40195327312566 (CS), Lot Number 25CBJ496; REF DYNJQ9041R: UDI/DI 10198459043086 (EA) 40198459043087 (CS), Lot Number 25FBF371; REF DYNJQ9041R: UDI/DI 10198459043086 (EA) 40198459043087 (CS), Lot Number 25EBK493; REF DYNJQ9041R: UDI/DI 10198459043086 (EA) 40198459043087 (CS), Lot Number 25DBH895; REF DYNJ900244I: UDI/DI 10195327464929 (EA) 40195327464920 (CS), Lot Number 25DBC272; REF DYNJ900244I: UDI/DI 10195327464929 (EA) 40195327464920 (CS), Lot Number 25CBU642; REF DYNJ900244I: UDI/DI 10195327464929 (EA) 40195327464920 (CS), Lot Number 25CBJ298; REF DYNJ900244I: UDI/DI 10195327464929 (EA) 40195327464920 (CS), Lot Number 25BBU206; REF DYNJ901281I: UDI/DI 10195327699154 (EA) 40195327699155 (CS), Lot Number 25FBA261; REF DYNJ901281I: UDI/DI 10195327699154 (EA) 40195327699155 (CS), Lot Number 25CBN230; REF DYNJ901829T: UDI/DI 10198459231384 (EA) 40198
Distribution
US Nationwide distribution.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-1089-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.