FDA Medical devices recall ·

Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS recall

Recall number
Z-1085-2026
Date
(initiated December 16, 2025)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
96 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC KIT-LF, REF DYNJ911359

Why was it recalled?

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

1) UDI/DI 10198459269196 (EA) 40198459269197 (CS), Lot Number 25EBD732; 2) UDI/DI 10198459326318 (EA) 40198459326319 (CS), Lot Number 25FBA103.

Distribution

US Nationwide distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1085-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.