FDA Medical devices recall ·
Access hsTnI Reagent recall
- Recall number
- Z-0727-2025
- Date
- (initiated November 18, 2024)
- Company / brand
- Beckman Coulter, Inc.
1000 Lake Hazeltine Dr, Chaska, MN, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 2,109,804 units
- Hazard tags
- Microbial contamination
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Access hsTnI Reagent, REF B52699, Immunoassay Method, Troponin Subunit. For the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma.
Why was it recalled?
Beckman Coulter, Inc. is recalling their Access hsTnI assay, a medical device by correction. The reason for the recall is: high troponin antigen (>55,000 pg/mL) can contaminate an analyzers probe and cause inter-assay carryover due to insufficient washing. The contamination can cause a falsely elevated Access hsTnI result in subsequent sample(s) which can impact patient care (potentially but not limited to unnecessary coronary imaging or diagnostic catheterization) if the result is near the medical decision points.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI/DI 15099590693183, All Lots
Distribution
Worldwide
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0727-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.