FDA Medical devices recall ·

ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in… recall

Recall number
Z-0202-2027
Date
(initiated September 21, 2026)
Company / brand
Beckman Coulter, Inc.
5355 W 76th St, Indianapolis, IN, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
336 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)

Why was it recalled?

Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Serial Numbers: 24U15250 to 26U15766; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860

Distribution

US Nationwide distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0202-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.