FDA Medical devices recall ·
Operating table column TS7500 MOBIUS, Hybrid operating table column, Model recall
- Recall number
- Z-0351-2027
- Date
- (initiated September 29, 2026)
- Company / brand
- Baxter Healthcare Corporation
1 Baxter Pkwy, Deerfield, IL, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 77 units
- Hazard tags
- Injury risk
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695
Why was it recalled?
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Serial numbers: 108094281, 108468075, 108484417, 108506318, 108516636, 108536785, 108544115, 108604154, 108463223, 108484404, 108489634, 108511411, 108516637, 108544106, 108544116, 108618964, 108463224, 108484405, 108489648, 108511412, 108521436, 108544107, 108604137, 108618965, 108468062, 108484406, 108506316, 108511413, 108521452, 108544110, 108604140, 108468063, 108484408, 108506317, 108516634, 108536776, 108544111, 108604153; UDI-DI number: 00887761974425
Distribution
Worldwide - US Nationwide distribution.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0351-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.