TruSystem 7000 U Operating Table (dV), Product Code 1723633 recall
- Recall number
- Z-0332-2027
- Date
- (initiated September 14, 2026)
- Company / brand
- Baxter Healthcare Corporation
25212 W Il Route 120, Round Lake, IL, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 62 units
- Hazard tags
- Injury risk
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
TruSystem 7000 U Operating Table (dV), Product Code 1723633
Why was it recalled?
The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI-DI: 00887761967519; Serial Number: 110188921, 110188922, and 110206662
Distribution
US Nationwide distribution in the states of IL, MO, TX.
States: Illinois Missouri Texas
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0332-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.