FDA Medical devices recall ·

Mobile column TruSystem 7500 U, Product Code 1730720 recall

Recall number
Z-0333-2027
Date
(initiated September 14, 2026)
Company / brand
Baxter Healthcare Corporation
25212 W Il Route 120, Round Lake, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1 unit
Hazard tags
Injury risk
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Mobile column TruSystem 7500 U, Product Code 1730720

Why was it recalled?

The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI: 00887761974388; Serial number: 110183858

Distribution

US Nationwide distribution in the states of IL, MO, TX.

States: Illinois Missouri Texas