FDA Medical devices recall ·

iFuse TORQ Implant System Model recall

Recall number
Z-0146-2027
Date
(initiated August 21, 2026)
Company / brand
SI-BONE, Inc.
471 El Camino Real Ste 101, Santa Clara, CA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
100 units
Hazard tags
Sterility
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

iFuse TORQ Implant System Model/Catalog Number: 10055T(10x55mm), 10060T(10x60mm), 10065T(10x65mm), 10080T(10x55mm), 11580T(11.5x80mm)

Why was it recalled?

The pelvic fixation screws failed to meet acceptance criteria for tube packaging seal integrity testing; therefore, sterility assurance cannot be ensured.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

10055T-UDI:00810055520077/Lot:9142871 10060T-UDI:00810055520084/Lot:9137761 10065T-UDI:00810055520091/Lot:9139213 10080T-UDI:00810055520121/Lot:9137781 11580T-UDI:00810055520367/Lot:9142911

Distribution

US Nationwide distribution in the states of FL, GA, HI, NC, OH, PA, SD, TX, VA.

States: Florida Georgia Hawaii North Carolina Ohio Pennsylvania South Dakota Texas Virginia

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.