FDA Medical devices recall ·

SmartPath to dStream for 1.5T, SmartPath to dStream for 3.0T… recall

Recall number
Z-3314-2026
Date
(initiated August 28, 2026)
Company / brand
Philips North America Llc
222 Jacobs St, Cambridge, MA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
200
Hazard tags
Software
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

SmartPath to dStream for 1.5T, SmartPath to dStream for 3.0T, SmartPath to dStream for XR and 3.0T, and SmartPath to Ingenia Elition X; Model number (REF): 1. 781260 (1.5T), 2. 782112 (1.5T), 3. 782146 (1.5T), 4. 782165 (1.5T), 5. 782145 (3.0T), 6. 782164 (3.0T), 7. 781270 (XR and 3.0T), 8. 782113 (XR and 3.0T), 9. 782129 (XR and 3.0T), 10. 782118 (Ingenia Elition X), 11. 782144 (Ingenia Elition X), 12. 782163 (Ingenia Elition X); Software Versions: R11, R12 and R13;

Why was it recalled?

When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

1. Model number (REF): 781260 (1.5T); UDI-DI: 00884838095076; Serial Numbers: 5479, 5490, 8057, 8118, 8260, 8330, 8392, 8819, 8884, 8954, 10044, 10170, 10236, 11062, 13018, 13029, 13047, 18663, 18872, 20033, 20115, 20164, 20319, 20322, 20413, 21011, 21062, 21133, 21142, 21156, 21278, 21577, 21583, 21621, 21775, 21880, 21969, 22157, 22248, 32049, 32245, 32262, 32286, 32322, 32845, 39024; 2. Model Number (REF): 782112 (1.5T); UDI-DI: 00884838098886; Serial Numbers: 12002, 21205, 21658, 30031, 32378, 32519, 33338; 3. Model Number (REF): 782146 (1.5T); UDI-DI: 00884838108691; Serial Numbers: 8935, 13013, 21202, 30308, 32570, 32634, 32679, 32683, 32750, 33074, 33479, 33611, 33714, 39040; 4. Model Number (REF): 782165 (1.5T); UDI-DI: 00884838115767; Serial Numbers: 21763, 22170, 32402, 32409, 32971, 33379, 33387, 33571, 39075, 72160, 72190; 5. Model Number (REF): 782145 (3.0T); UDI-DI: 00884838108684; Serial Numbers: 17377, 17385, 17435, 24010, 24062, 34256, 34282, 38059, 38118, 38127, 38170, 38251, 38278, 38294, 38295, 38302, 38336, 73002, 74004, 74007, 74030; 6. Model Number (REF): 782164 (3.0T); UDI-DI: 00884838115750; Serial Numbers: 34038, 34254, 34268, 38021, 38046, 38061, 38086, 38090, 38223, 38246, 74020, 74033; 7. Model Number (REF): 781270 (XR and 3.0T); UDI-DI: 00884838095083; Serial Numbers: 17239, 17246, 17264, 17269, 17283, 17299, 17356, 17357, 17369, 17406, 17416, 17425, 17434, 17464, 17468, 17473, 17514, 17544, 17550, 17564, 17565, 17569, 17589, 17590, 23026, 23033,

Distribution

Worldwide distribution - Nationwide and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Equator. Guinea, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kyrgyzstan, Latvia, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, T¿rkiye, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam.

States: Georgia

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-3314-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.