FDA Medical devices recall ·

Philips Medical Systems Technologies Ltd recall

Recall number
Z-3155-2026
Date
(initiated August 18, 2026)
Company / brand
Philips North America Llc
222 Jacobs St, Cambridge, MA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
62
Hazard tags
Crash risk
Initiated by
FDA Mandated

Read the official FDA notice ↗

Product

Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333

Why was it recalled?

The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI: (01)00884838101111

Distribution

U.S. and U.S. Territories.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-3155-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.