Philips Medical Systems Technologies Ltd recall
- Recall number
- Z-3155-2026
- Date
- (initiated August 18, 2026)
- Company / brand
- Philips North America Llc
222 Jacobs St, Cambridge, MA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 62
- Hazard tags
- Crash risk
- Initiated by
- FDA Mandated
Read the official FDA notice ↗
Product
Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333
Why was it recalled?
The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.