FDA Medical devices recall ·

Adapter for Head Positioning, Model recall

Recall number
Z-2638-2026
Date
(initiated April 21, 2026)
Company / brand
Baxter Healthcare Corporation
1 Baxter Pkwy, Deerfield, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
13 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Adapter for Head Positioning, Model/Catalog Number: 1739994, when used in conjunction with the side rails of the Upper Back Section H V U component (product code 1909813) on Baxter operating tables

Why was it recalled?

The Adapter for Head Positioning accessory may become loose during patient positioning and unintentionally move when used in conjunction with the side rails of the Upper Back Section H V U component on Baxter operating tables.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Lot Code: Product Code: 1739994 UDI-DI: 00887761972780 Serial Numbers: 109931797, 109931798, 109931799, 109943588, 109943590, 109948550, 109964963, 109964964, 109964969, 109964970, 109964971, 109980700, and 110006694

Distribution

Worldwide - US Nationwide distribution in the states of Minnesota, Maryland, and Pennsylvania and the countries of Canada, Hong Kong, India, Japan, Korea, Philippines, Taiwan, Thailand, Vietnam, Austria, Belgium, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Iraq, Israel, Italy, Jordan, Kenya, Lebanon, Libya, Malta, Morocco, Netherlands, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Romania, Saudi Arabia, Senegal, Serbia, Slovakia, Spain, Sweden, Switzerland, Tunisia, Turkey, United Arabia Emrites, Ukraine, Argentina, Brazil, Chile, Colombia, Costa Rica, Mexico, and Peru.

States: Maryland Minnesota Pennsylvania

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2638-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.