FDA Medical devices recall ·
Medtronic Harmony Delivery Catheter System recall
This is an FDA Class I recall — the most serious type.
- Recall number
- Z-2624-2026
- Date
- (initiated May 22, 2026)
- Company / brand
- Medtronic Heart Valves Division
1851 E Deere Ave, Santa Ana, CA, United States - Category
- Medical devices › Class I
- Classification
- Class I
- Status
- Ongoing
- Units affected
- 1881 units
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Medtronic Harmony Delivery Catheter System
Why was it recalled?
Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a manufacturing issue.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
GTIN 00763000341367, Lot Numbers: 0012508285, 0012517045, 0012526102, 0012526103, 0012614786, 0012614787, 0012623740, 0012623741, 0012642959, 0012659419, 0012663878, 0012663879, 0012690218, 0012699635, 0012702022, 0012710787, 0012728367, 0012728369, 0012777713, 0012804152, 0012804579, 0012824770, 0012824771, 0012874526, 0012896853, 0012896884, 0012910323, 0012965210, 0012965211, 0012976400, 0012986693, 0013063766, 0013063767, 0013105039, 0013105040, 0013133457, 0013181623, 0013181624, 0013236643, 0013272796, 0013272798. GTIN 00763000544027, Lot Numbers: 0012508285, 0012526103, 0012614786, 0012614787, 0012623740, 0012623741, 0012642959, 0012659419, 0012663879, 0012702022, 0012710787, 0012728367, 0012728369, 0012777713, 0012804152, 0012804579, 0012824771, 0012874526, 0012896853, 0012896884, 0012910323, 0012965211, 0012976400, 0012986693, 0013063767, 0013105039, 0013105040, 0013133457, 0013181623, 0013181624. GTIN 00763000520151, Lot Numbers: 0012575897, 0012642965, 0012728368, 0013272799. GTIN 00763000582951, Lot Numbers: 0012590535, 0012606411, 0012606413, 0012642964, 0012659418, 0012678282, 0012690219, 0012699637, 0012804151, 0012804580, 0012874527, 0012888682, 0012896885, 0012965221, 0012986699, 0013063764, 0013063765, 0013105037, 0013227504. GTIN 00763000918712, Lot Numbers: 0012635392, 0012676200, 0012690207, 0012874528, 0012874529, 0013015539, 0013105038.
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Canada, Chile, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Mexico, Netherlands, Northern Ireland, Norway, Oman, Poland, Portugal, Puerto Rico, Qatar, Romania, Saudi Arabia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
States: Puerto Rico
What does Class I mean?
Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.
Source: FDA (medical devices), recall Z-2624-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.