FDA Medical devices recall ·

Omnipod (Eros) Pod recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-2605-2026
Date
(initiated May 20, 2026)
Company / brand
Insulet Corporation
50 & 100 Nagog Park, Acton, MA, United States
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
289,046 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Omnipod (Eros) Pod. Includes the below Model/REF Numbers: 1. POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). 2. ZXR425 (10-Pack), ZXR420 (Single Pod). 3. ZXP425 (10 Pack), ZXP420 (Single Pod).

Why was it recalled?

External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

1. Model/REF Number: POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10385081120036. Lot Numbers: L72514. 2. Model/REF Number: ZXR425 (10-Pack), ZXR420 (Single Pod). Pack UDI-DI: N/A. Single Pod UDI-DI: N/A. Lot Numbers: L72516, L72521. 3. Model/REF Number: ZXP425 (10 Pack), ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10385081120036. Lot Numbers: L72515, L72509, L72513, L72511, L72510, L71480, L72514.

Distribution

Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Switzerland, Germany, Denmark, Finland, France, UK, Greece, Israel, Italy, Kuwait, Leichtenstein, Netherlands, Norway, Saudi Arabia, Sweden and Turkey.

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-2605-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.