FDA Medical devices recall ·

UniCel DxH 800 COULTER Cellular Analysis System… recall

Recall number
Z-2574-2025
Date
(initiated August 13, 2025)
Company / brand
Beckman Coulter, Inc.
250 S Kraemer Blvd, Brea, CA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
N/A
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

UniCel DxH 800 COULTER Cellular Analysis System, configured with the Led HGB Photometer REFs:B24802, 629029, B63322

Why was it recalled?

Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDIs: 15099590672423, 15099590381646, 15099590687922/ Serial Numbers: AT38527, AU02008, AU10078, AV28630, AW13187, AW29123, BC18100, RBB03201, AW43276, AY06074, AY06092, AY41556, AY43605, AZ02030, AZ03041, AZ18245, AZ22329, AZ39513, AZ46681, BA03048, BA03050, BA03055, BA03060, BA07061, BA07062, BA13227, BA15245, BA27414, BB16181, BB29286, BC08036, BC18099, AZ25376, BB25262, BB25261, BA25393

Distribution

Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Algeria, Argentina, Australia, Bahrain, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Egypt, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Jordan, Kenya, Korea Republic of, Kuwait, Kyrgyzstan, Libya, Lithuania, Mexico, Nepal, Netherlands, New Zealand, North Macedonia, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Switzerland, Taiwan, Province of China, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, United States of America, Uruguay, Vietnam.

States: Alabama Arizona California Connecticut Florida Georgia Iowa Idaho Illinois Indiana Kentucky Louisiana Massachusetts Maryland Michigan Minnesota Missouri Mississippi Montana North Carolina Nebraska New Jersey New York Ohio Oklahoma Oregon Pennsylvania Puerto Rico Rhode Island South Carolina South Dakota Tennessee Texas Virginia Washington Wisconsin West Virginia

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.