FDA Medical devices recall ·

DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457 recall

Recall number
Z-2267-2025
Date
(initiated June 25, 2025)
Company / brand
Beckman Coulter, Inc.
1000 Lake Hazeltine Dr, Chaska, MN, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1 system
Hazard tags
Software
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457

Why was it recalled?

Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI: 15099590369231; Serial Numbers: 608109

Distribution

Worldwide distribution - US Nationwide.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2267-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.