FDA Medical devices recall ·

REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US)… recall

Recall number
Z-2888-2026
Date
(initiated June 26, 2026)
Company / brand
Beckman Coulter, Inc.
250 S Kraemer Blvd, Brea, CA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
158 instruments
Hazard tags
Microbial contamination
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284 MicroScan WalkAway-96 plus Instrument - B1018-440 DxM-1040 MicroScan WalkAway Instrument - B1018-496 DxM-1096 MicroScan WalkAway Instrument

Why was it recalled?

Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI: 15099590799243/ V5.03 Panel Update 11

Distribution

US Nationwide distribution including in the states of AK, AL, AR, CA, CO, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WI, WV, WY.

States: Alaska Alabama Arkansas California Colorado Delaware Florida Georgia Iowa Illinois Indiana Kentucky Louisiana Massachusetts Maryland Maine Michigan Minnesota Missouri Mississippi North Carolina North Dakota Nebraska New Jersey Nevada New York Ohio Oklahoma Oregon Pennsylvania South Carolina Tennessee Texas Virginia Wisconsin West Virginia Wyoming

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2888-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.