FDA Medical devices recall ·

Pediatric care bed recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-2212-2026
Date
(initiated May 4, 2026)
Company / brand
Kayserbetten Gmbh & CO. Kg
Rieper Str. 12, Bad Fallingbostel, Germany
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
372 units
Hazard tags
Injury risk
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Pediatric care bed; Product Designation: KayserBett IDA;

Why was it recalled?

If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap themselves or others under the bed frame or between the bed frame and the floor. This can lead to serious injuries or even death of the patient (child) or third parties (children).

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Product Designation: KayserBett IDA; UDI-DI: 426038961IDAGU; Serial Number: 386, 387, 388, 389, 390, 391, 392, 635, 636, 637, 638, 639, 641, 642, 643, 644, 645, 646, 647, 1072, 1073, 1074, 1075, 1076, 1077, 1078, 1079, 1080, 1081, 1082, 1083, 1085, 1086, 1087, 1088, 1089, 1090, 1110, 1111, 1446, 1447, 1448, 1449, 1648, 1649, 1950, 1975, 1976, 1982, 1983, 1998, 2001, 2003, 2004, 2337, 2338, 2339, 2340, 2341, 2342, 2343, 2344, 2345, 2396, 2763, 2764, 2765, 2766, 2767, 2770, 2771, 2772, 2773, 2774, 3087, 3088, 3089, 3090, 3091, 3092, 3167, 3299, 3300, 3301, 3302, 3303, 3304, 3305, 3306, 3309, 3310, 3311, 3312, 3313, 3314, 3740, 3741, 3760, 4223, 4232, 4233, 4234, 4235, 4236, 4237, 4239, 4240, 4242, 4243, 4244, 4245, 4246, 4247, 4248, 4249, 4250, 4567, 4568, 4569, 4570, 4571, 4572, 4573, 4574, 4575, 4576, 4577, 4578, 4579, 4580, 4581, 4582, 4583, 4584, 4986, 5238, 5254, 5255, 5256, 5257, 5258, 5259, 5637, 5919, 5920, 5921, 5922, 5923, 5924, 5925, 5926, 5937, 5938, 5939, 5940, 5948, 6332, 6338, 6339, 6340, 6341, 6342, 6343, 6344, 6345, 6346, 6347, 6351, 6355, 6795, 6796, 7176, 7177, 7178, 7179, 7180, 7181, 7182, 7183, 7184, 7185, 7186, 7187, 7188, 7189, 7190, 7191, 7192, 7193, 7194, 7195, 7617, 7618, 7619, 7620, 7621, 7622, 7623, 8242, 8243, 8244, 8245, 8588, 8589, 8590, 8591, 8592, 8593, 8594, 8595, 8596, 8597, 8598, 8599, 8600, 8601, 8602, 8603, 8608, 8609, 8691, 8704, 8912, 8921, 8922, 8923, 8924, 8925, 8926, 8927, 8928, 8929, 8930, 9725, 9726, 9727, 10350, 10351, 10352, 10819,

Distribution

Distribution US Nationwide and Canada.

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-2212-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.