- Kayserbetten Gmbh & CO. Kg
Rieper Str. 12, Bad Fallingbostel, Germany - Category
- Medical devices › Class I
- Classification
- Class I
- Status
- Ongoing
- Units affected
- 372 units
- Hazard tags
- Injury risk
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Pediatric care bed; Product Designation: KayserBett IDA;
Why was it recalled?
If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap themselves or others under the bed frame or between the bed frame and the floor. This can lead to serious injuries or even death of the patient (child) or third parties (children).
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Product Designation: KayserBett IDA; UDI-DI: 426038961IDAGU; Serial Number: 386, 387, 388, 389, 390, 391, 392, 635, 636, 637, 638, 639, 641, 642, 643, 644, 645, 646, 647, 1072, 1073, 1074, 1075, 1076, 1077, 1078, 1079, 1080, 1081, 1082, 1083, 1085, 1086, 1087, 1088, 1089, 1090, 1110, 1111, 1446, 1447, 1448, 1449, 1648, 1649, 1950, 1975, 1976, 1982, 1983, 1998, 2001, 2003, 2004, 2337, 2338, 2339, 2340, 2341, 2342, 2343, 2344, 2345, 2396, 2763, 2764, 2765, 2766, 2767, 2770, 2771, 2772, 2773, 2774, 3087, 3088, 3089, 3090, 3091, 3092, 3167, 3299, 3300, 3301, 3302, 3303, 3304, 3305, 3306, 3309, 3310, 3311, 3312, 3313, 3314, 3740, 3741, 3760, 4223, 4232, 4233, 4234, 4235, 4236, 4237, 4239, 4240, 4242, 4243, 4244, 4245, 4246, 4247, 4248, 4249, 4250, 4567, 4568, 4569, 4570, 4571, 4572, 4573, 4574, 4575, 4576, 4577, 4578, 4579, 4580, 4581, 4582, 4583, 4584, 4986, 5238, 5254, 5255, 5256, 5257, 5258, 5259, 5637, 5919, 5920, 5921, 5922, 5923, 5924, 5925, 5926, 5937, 5938, 5939, 5940, 5948, 6332, 6338, 6339, 6340, 6341, 6342, 6343, 6344, 6345, 6346, 6347, 6351, 6355, 6795, 6796, 7176, 7177, 7178, 7179, 7180, 7181, 7182, 7183, 7184, 7185, 7186, 7187, 7188, 7189, 7190, 7191, 7192, 7193, 7194, 7195, 7617, 7618, 7619, 7620, 7621, 7622, 7623, 8242, 8243, 8244, 8245, 8588, 8589, 8590, 8591, 8592, 8593, 8594, 8595, 8596, 8597, 8598, 8599, 8600, 8601, 8602, 8603, 8608, 8609, 8691, 8704, 8912, 8921, 8922, 8923, 8924, 8925, 8926, 8927, 8928, 8929, 8930, 9725, 9726, 9727, 10350, 10351, 10352, 10819,
Distribution
Distribution US Nationwide and Canada.
What does Class I mean?
Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.