FDA Medical devices recall ·

RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers:… recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-2160-2026
Date
(initiated April 22, 2026)
Company / brand
Bolton Medical Inc.
799 International Pkwy, Sunrise, FL, United States
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-34-109-34U 28-N4-34-154-30U 28-N4-34-154-34U 28-N4-34-209-30U 28-N4-34-209-34U 28-N4-34-259-34U 28-N4-36-109-36U 28-N4-36-154-32U 28-N4-36-154-36U 28-N4-36-199-32U 28-N4-36-199-36U 28-N4-36-259-36U 28-N4-38-109-38U 28-N4-38-154-34U 28-N4-38-154-38U 28-N4-38-199-34U 28-N4-38-199-38U 28-N4-38-259-38U 28-N4-40-114-40U 28-N4-40-154-36U 28-N4-40-154-40U 28-N4-40-204-36U 28-N4-40-204-40U 28-N4-40-259-40U 28-N4-42-114-42U 28-N4-42-159-38U 28-N4-42-159-42U 28-N4-42-204-38U 28-N4-42-204-42U 28-N4-42-259-42U 28-N4-44-114-44U 28-N4-44-164-40U 28-N4-44-164-44U 28-N4-44-209-40U 28-N4-44-209-44U 28-N4-44-259-40U 28-N4-44-259-44U 28-N4-46-114-46U 28-N4-46-164-42U 28-N4-46-164-46U 28-N4-46-209-42U 28-N4-46-209-46U 28-N4-46-259-42U

Why was it recalled?

Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

All Lots/(Ref Number/Proximal Diameter/DI)28-N4-32-104-32U/ 32/ (01)00843576150461; 28-N4-32-164-28U/ 32/ (01)00843576150959; 28-N4-32-164-32U/ 32/ (01)00843576150591; 28-N4-32-209-28U/ 32/ (01)00843576151055; 28-N4-32-209-32U/ 32/ (01)00843576150720; 28-N4-32-259-32U/ 32/ (01)00843576150850; 28-N4-34-109-34U/ 34/ (01)00843576150478; 28-N4-34-154-30U/ 34/ (01)00843576150966; 28-N4-34-154-34U/ 34/ (01)00843576150607; 28-N4-34-209-30U/ 34/ (01)00843576151062; 28-N4-34-209-34U/ 34/ (01)00843576150737; 28-N4-34-259-34U/ 34/ (01)00843576150867; 28-N4-36-109-36U/ 36/ (01)00843576150485; 28-N4-36-154-32U/ 36/ (01)00843576150973; 28-N4-36-154-36U/ 36/ (01)00843576150614; 28-N4-36-199-32U/ 36/ (01)00843576151079; 28-N4-36-199-36U/ 36/ (01)00843576150744; 28-N4-36-259-36U/ 36/ (01)00843576150874; 28-N4-38-109-38U/ 38/ (01)00843576150492; 28-N4-38-154-34U/ 38/ (01)00843576150980; 28-N4-38-154-38U/ 38/ (01)00843576150621; 28-N4-38-199-34U/ 38/ (01)00843576151086; 28-N4-38-199-38U/ 38/ (01)00843576150751; 28-N4-38-259-38U/ 38/ (01)00843576150881; 28-N4-40-114-40U/ 40/ (01)00843576150508; 28-N4-40-154-36U/ 40/ (01)00843576150997; 28-N4-40-154-40U/ 40/ (01)00843576150638; 28-N4-40-204-36U/ 40/ (01)00843576151093; 28-N4-40-204-40U/ 40/ (01)00843576150768; 28-N4-40-259-40U/ 40/ (01)00843576150898; 28-N4-42-114-42U/ 42/ (01)00843576150515; 28-N4-42-159-38U/ 42/ (01)00843576151000; 28-N4-42-159-42U/ 42/ (01)00843576150645; 28-N4-42-204-38U/ 42/ (01)00843576151109; 28-N4-42-204-42U/ 42/ (01)0084

Distribution

Worldwide - distribution in the US Territory of Puerto Rico and the countries of Argentina, Chile, Canada, Colombia, Mexico, United Kingdom.

States: Puerto Rico

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-2160-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.