FDA Medical devices recall ·

Hillrom LikoScale component, Product Code 3156232… recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-2032-2025
Date
(initiated May 30, 2025)
Company / brand
Baxter Healthcare Corporation
1 Baxter Pkwy, Deerfield, IL, United States
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
9839 units
Hazard tags
Fall Injury risk
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.

Why was it recalled?

The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

All LikoScale Adapter Kits manufactured between 8/27/2013 to 2/27/2025

Distribution

Worldwide

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-2032-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.