FDA Medical devices recall ·

Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Model recall

Recall number
Z-3215-2026
Date
(initiated August 6, 2026)
Company / brand
Heidelberg Engineering GmbH
Max-Jarecki-Str. 8, Heidelberg, Germany
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
87
Hazard tags
Fall Tip-over Injury risk
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Model/Catalog Number: S3610, S2610, S2000, S1610 multimodal imaging platform optimized for the posterior segment of the eye

Why was it recalled?

Wheels may come loose when module is being pushed or pulled during movement or transportation, affecting stability of the stand. The stand may tip over as a result. Falling stand may cause impact or crushing injuries, particularly to the feet, legs, hands or arms of the operator or nearby persons.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI 04059237000235 Lot Code: Spec-07265, Spec-11121, Spec-11281, Spec-12096, Spec-13305, Spec-14403, Spec-15282, Spec-15587, Spec-20071, Spec-20348, Spec-20351, Spec-20591, Spec-20605, Spec-23473, Spec-23941, Spec-23950, Spec-23975, Spec-24507, Spec-24558, Spec-24582, Spec-24611

Distribution

Worldwide - US Nationwide distribution in the state of Massachusetts and the countries of Australia, Costa Rica, Germany, France, India, Israel, Italy, Canada, Mexico, Netherlands, Austria, Philippines, Poland, Saudi Arabia, Switzerland, Spain. South Africa. Thailand, Czech Republic, United Kingdom, Vietnam.

States: Massachusetts

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-3215-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.