FDA Medical devices recall ·
Newport HT70 Ventilator recall
This is an FDA Class I recall — the most serious type.
- Recall number
- Z-1915-2025
- Date
- (initiated May 21, 2025)
- Company / brand
- Covidien, LP
2101 Faraday Ave, Carlsbad, CA, United States - Category
- Medical devices › Class I
- Classification
- Class I
- Status
- Ongoing
- Units affected
- 5394
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Newport HT70 Ventilator, REF: HT70M-JP-NA, HT70M-CN-NA, HT70M-ES-EU, HT70M-PT-BR, HT70M-SY-EU, HT70M-WW-EU, HT70M-WW-NA, HT70-SY-EU. Newport HT70Plus Ventilator, REF: HT70PM-ES-EU, HT70PM-ES-NA, HT70PM-JP-NA, HT70PM-PT-BR, HT70PM-SY-AS, DLHT70PM-WW-NA, HT70PM-SY-NA, HT70PM-SY-UK, HT70PM-WW-EU, HT70PM-WW-UK, HT70PM-SY-EU, HT70PM-WW-NA. HT70-2 CONTROL BOARD ROHS, REF: GR-PCB3252A. HT70-2 CONTROL BOARD ROHS X1, REF: GR105796. HT70P CONTROL BOARD ROHS X1, REF: GR105795. HT70 Conversion Kit, REF: GRFSKHT70M-2.
Why was it recalled?
Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to alarm effectively during shut down, which may result in respiratory failure, hypoventilation, low oxygen saturation, hypoxia, treatment delay.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
REF/UDI-DI/Serial Numbers: HT70M-JP-NA/00839935008221/N10HT700710132, N10HT700710135, N10HT700710136, N10HT700710159, N10HT700710161, N10HT700710162, N10HT700710163, N10HT700710164, N10HT700810194, N10HT700810199, N10HT700810202, N10HT700810204, N10HT700810205, N10HT700810206, N10HT700910223, N10HT700910227, N10HT701010279, N10HT701011015, N10HT701011016, N10HT701111044, N10HT701111048, N10HT701111049, N10HT701111050, N11HT700111137, N11HT700111139, N11HT700111141, N11HT700111142, N11HT700111143, N11HT700111144, N11HT700211145, N11HT700211155, N11HT700211160, N11HT700211161, N11HT700211163, N11HT700211164, N11HT700311168, N11HT700311169, N11HT700311170, N11HT700311183, N11HT700311184, N11HT700311185, N11HT700311186, N11HT700311188, N11HT700311189, N11HT700311191, N11HT700311192, N11HT700311193, N11HT700311195, N11HT700311196, N11HT700311197, N11HT700311201, N11HT700311202, N11HT700311203, N11HT700311204, N11HT700311207, N11HT700311208, N11HT700311209, N11HT700311233, N11HT700311234, N11HT700311243, N11HT700311244, N11HT700311245, N11HT700311246, N11HT700311247, N11HT700311248, N11HT700311250, N11HT700311251, N11HT700311253, N11HT700311255, N11HT700411256, N11HT700411257, N11HT700411258, N11HT700411262, N11HT700411263, N11HT700411264, N11HT700411265, N11HT700411266, N11HT700411267, N11HT700411268, N11HT700411271, N11HT700411277, N11HT700411278, N11HT700411279, N11HT700411282, N11HT700411284, N11HT700411285, N11HT700411286, N11HT700411287, N11HT700411288, N11HT700411289, N11HT7
Distribution
Worldwide - US Nationwide distribution including in the states of MA, CT, KS, OH, PR, NH, GA, WI, MS, VA, CO, TX, MN, AL, IA, NY, UT, NM, KY, RI, SD, PA, IA, CA, MO, HI, MI, IL, MT, NJ, TN, SC, FL, IN, AZ, ID, LA, WA, NC, MD, WY, NE, NV, ND, ME, AK, DE, OK, OR and the countries of Japan, Canada, Singapore, Lithuania, Kazakhstan, Slovenia, Colombia, China, Qatar, United Arab Emirates, Greece, Saudi Arabia, Nigeria, Romania, Netherlands, Lebanon, Oman, Jordan, Malawi, India, Croatia, Portugal, Turkey, Brazil, Pakistan, Trinidad and Tobago, Slovakia, Chile, Egypt, Switzerland, South Africa, Thailand, Bangladesh, Myanmar, Peru, Uruguay, Hungary, Argentina, Iran, Islamic Republic Of, Costa Rica, Vietnam, Australia, Guatemala, Mexico, Guyana, Jamaica, Dominican Republic, Bahamas, Bosnia and Herzegovina, South Korea, Tunisia, Ukraine, Uzbekistan, Indonesia, Burkina Faso, Italy, Panama, Sri Lanka, Malaysia, Namibia, Swaziland, Spain, Philippines, Bhutan, Mauritius, Poland, France, Cambodia, Czechia, Israel, Norway, Russian Federation, United Kingdom, Canary Islands, Kenya, Czech Republic, Croatia , Estonia, Maldova, Palestine.
States: Alaska Alabama Arizona California Colorado Connecticut Delaware Florida Georgia Hawaii Iowa Idaho Illinois Indiana Kansas Kentucky Louisiana Massachusetts Maryland Maine Michigan Minnesota Missouri Mississippi Montana North Carolina North Dakota Nebraska New Hampshire New Jersey New Mexico Nevada New York Ohio Oklahoma Oregon Pennsylvania Puerto Rico Rhode Island South Carolina South Dakota Tennessee Texas Utah Virginia Washington Wisconsin Wyoming
What does Class I mean?
Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.
Source: FDA (medical devices), recall Z-1915-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.