FDA Medical devices recall ·

Olympus EVIS EXERA III Duodenovideoscope recall

Recall number
Z-1911-2025
Date
(initiated May 5, 2025)
Company / brand
Olympus Corporation of the Americas
3500 Corporate Pkwy, Center Valley, PA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
12,195 units
Hazard tags
Microbial contamination
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190V

Why was it recalled?

Recent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed, compared to those TJF duodenoscopes where manual cleaning began within one hour after patient procedure.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI: 04953170405563. Model Number: TJF-Q190V. All serial numbers.

Distribution

Worldwide Distribution: US (Nationwide) and OUS (International) countries to: Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Benin, Bolivia, Brazil, Canada, Cayman Islands, Chile, Colombia, Costa Rica, Croatia, Denmark, Dominican Republic, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Finland, France, French Guiana, Georgia, Germany, Ghana, Guadeloupe, Guatemala, Honduras, Hong Kong, Iceland, India, Iraq, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Morocco, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Palestinian Authority, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Republic of the Congo, R¿union, Russia, Saudi Arabia, Senegal, Seychelles, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, Uruguay, Vietnam, Yemen.

States: Georgia

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1911-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.