FDA Medical devices recall ·

IceSeed 1.5 CX S NEEDLE US, Cryoablation Needle recall

Recall number
Z-1686-2025
Date
(initiated March 24, 2025)
Company / brand
Boston Scientific Corporation
1 Scimed Pl, Maple Grove, MN, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
435 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

IceSeed 1.5 CX S NEEDLE US, Cryoablation Needle, REF H7493967233100. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.

Why was it recalled?

Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

GTIN 00191506032678, Batch Numbers: 33744602, 33744603, 33744604, 33744605, 33931901, 33937786, 33937787, 33942987, 33942988, 33942989, 33950417, 33950418, 33996423, 33996424, 33996428, 34038291, 34038292, 34038307, 34117281, 34128141, 34128142, 34151877, 34151878, 34204928, 34204929, 34256930, 34256931, 34256932, 34295233.

Distribution

Nationwide

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1686-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.