FDA Medical devices recall ·

Medline Industries IV Administration and Extension Sets: 1) DBD-IV ADMIN… recall

Recall number
Z-1366-2026
Date
(initiated December 31, 2025)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
926,227 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medline Industries IV Administration and Extension Sets: 1) DBD-IV ADMIN SET,60DRP,1NEEDLE-FREE VALV, Model Number: DYNDTB0529; 2) IV ADMIN SET, 10DRP, 1NF, 92, Model Number: DYNDTB1029; 3) DBD-IVEXTSET2.4ML2 NEEDLE-FREEVLVE,35, Model Number: DYNDTB1516; 4) DBD-IV ADMIN SET, 15DRP, 1 INJECT PORT,, Model Number: DYNDTB1530; 5) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 6, Model Number: DYNDTB1539; 6) DBD-IV ADMIN SET, 15DRP, 3 NF, 109, Model Number: DYNDTB1545; 7) IV ADMIN SET, 20DRP, 3 NF, 109, Model Number: DYNDTB2045; 8) DBD-IVADMINSET60DRP2NEEDLE-FREEVLVE96, Model Number: DYNDTB0512; 9) DBD-IVADMINSET,60DRP3 NEEDLE-FREE VLVE11, Model Number: DYNDTB0555; 10) IV ADMIN SET, 10DRP, NF,INJ, 96, Model Number: DYNDTB1022; 11) DBD-IV ADMIN SET, 15DRP, 2NF, 96, Model Number: DYNDTB1512; 12) DBD-IV ADMIN SET, 15DRP, 1NF, 92, Model Number: DYNDTB1529; 13) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 9, Model Number: DYNDTB1537; 14) DBD-IV ADMIN SET, 15DRP, SECOND, 40, Model Number: DYNDTB1540; 15) IV ADMIN SET, 20DRP, 1NF, 92, Model Number: DYNDTB2029; 16) DBD-IV EXT SET, 6.5, 1.1ML, M LUERLOCK,, Model Number: DYNDTB5077; 17) IV ADMIN SET, 10DRP, 3 NF, 109, Model Number: DYNDTB1045; 18) DBD-IVADMISET15DRP4NEEDLEFREE VLVESTPCK1, Model Number: DYNDTB1545D; 19) IV ADMIN SET15DRPFLWCTRL1NEEDLEFREE VA, Model Number: DYNDTB2524; 20) IV ADMIN SET, 10DRP, 2NF, 96, Model Number: DYNDTB1012; 21) IV ADMIN SET, 20DRP, 2NF, 96, Model Number: DYNDTB2012; 22) IV ADMIN SET, 20DRP, NF,INJ, 96, Model Number: DYNDTB2022; 23) DBD-IV EXT SET,SMBORE EXT,8,M LUERLOCK,1, Model Number: DYNDTB5081

Why was it recalled?

Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may result in a delay in patient treatment, blood loss, infection, an air embolism, and/or underdosage of medication.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

1) DYNDTB0529, UDI-DI: 10197344105663(each), 20197344105660(case), Lot Number: All Lot Numbers; 2) DYNDTB1029, UDI-DI: 10197344076956(each), 20197344076953(case), Lot Number: All Lot Numbers; 3) DYNDTB1516, UDI-DI: 10197344105687(each), 20197344105684(case), Lot Number: All Lot Numbers; 4) DYNDTB1530, UDI-DI: 10197344105694(each), 20197344105691(case), Lot Number: All Lot Numbers; 5) DYNDTB1539, UDI-DI: 10197344008254(each), 20197344008251(case), Lot Number: All Lot Numbers; 6) DYNDTB1545, UDI-DI: 10197344098644(each), 20197344098641(case), Lot Number: All Lot Numbers; 7) DYNDTB2045, UDI-DI: 10197344077007(each), 20197344077004(case), Lot Number: All Lot Numbers; 8) DYNDTB0512, UDI-DI: 10197344105656(each), 20197344105653(case), Lot Number: All Lot Numbers; 9) DYNDTB0555, UDI-DI: 10197344105670(each), 20197344105677(case), Lot Number: All Lot Numbers; 10) DYNDTB1022, UDI-DI: 10197344076949(each), 20197344076946(case), Lot Number: All Lot Numbers; 11) DYNDTB1512, UDI-DI: 10197344098613(each), 20197344098610(case), Lot Number: All Lot Numbers; 12) DYNDTB1529, UDI-DI: 10197344098620(each), 20197344098627(case), Lot Number: All Lot Numbers; 13) DYNDTB1537, UDI-DI: 10197344008247(each), 20197344008244(case), Lot Number: All Lot Numbers; 14) DYNDTB1540, UDI-DI: 10197344098637(each), 20197344098634(case), Lot Number: All Lot Numbers; 15) DYNDTB2029, UDI-DI: 10197344076994(each), 20197344076991(case), Lot Number: All Lot Numbers; 16) DYNDTB5077, UDI-DI: 10197344098651(each), 20197344

Distribution

Worldwide distribution - US Nationwide and the country of Bahamas.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1366-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.