FDA Medical devices recall ·

Access Erythropoietin (EPO) recall

Recall number
Z-1249-2025
Date
(initiated February 3, 2025)
Company / brand
Beckman Coulter, Inc.
1000 Lake Hazeltine Dr, Chaska, MN, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
498 US; 1576 OUS
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 Daltons) produced primarily by the kidney, is the principal factor regulating red blood cell production (erythropoiesis) in mammals. Renal production of EPO is regulated by changes in oxygen availability. Under conditions of hypoxia, the level of EPO in the circulation increases and this leads to increased production of red blood cells.

Why was it recalled?

Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the affected lot may demonstrate repeatable falsely decreased results, which may lead to improper diagnosis or repeat testing.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Catalog A16364 UDI-DI 15099590201838 Lot 439363

Distribution

Domestic distribution to AL AR AZ CA CT FL ID IL IN KS MA MI NC NE NJ NY OH OR PA TN TX UT WA WV. International distribution to Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Czechia, Denmark, France, Germany, Hungary, India, Italy, Japan, Korea, Republic of, Netherlands, Philippines, Russian Federation, Saudi Arabia, Slovakia, Spain, United Kingdom of Great Britain and Northern Ireland

States: Arkansas Arizona California Connecticut Florida Illinois Kansas Michigan North Carolina Nebraska New Jersey New York Tennessee Texas Utah Washington West Virginia

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1249-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.