FDA Medical devices recall ·

Medtronic Pipeline Vantage Embolization Device with Shield Technology recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-1246-2025
Date
(initiated January 30, 2025)
Company / brand
Micro Therapeutics, Inc.
9775 Toledo Way, Irvine, CA, United States
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
7820
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PED3-021-275-12, PED3-021-275-16, PED3-021-325-12, PED3-021-325-14, PED3-021-325-16, PED3-021-325-20, PED3-021-250-12, PED3-021-300-14, PED3-021-250-16, PED3-021-300-12, PED3-021-275-14, PED3-021-350-14, PED3-021-350-25, PED3-021-350-12, PED3-021-250-10, PED3-021-275-20, PED3-021-350-18, PED3-021-325-18, PED3-021-275-18, PED3-021-250-18, PED3-021-300-18; and Instructions for Use, Part: M056989CDOC2

Why was it recalled?

Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Instructions for Use, Version: Rev. A (01/2024). REF/UDI-DI/Lot: PED3-021-350-16/00847536030701, 00847536035195, 00847536036529/B317460, B327449, B335184, B340948, B349032, B389162, B392918, B424454, B430015, B495821, B497938, B547396, B550307, B551615, B559033, B603377, B623564, B656863, B656887, B658904, B692346, B715057, B715058, B715541, B726894, B735821, B735823, B756029, B756030, B760210, B760211, B768202, B768387, B769453, B770633, B807494; PED3-021-250-14/00847536030411, 00847536034907, 00847536036239/B317936 B332767, B333492, B333536, B334545, B376967, B388041, B391185, B399354, B416713, B431395, B452416, B457963, B467691, B476239, B564173, B564970, B574472, B580923, B629372, B630998, B670305, B673035, B699576, B707568, B709193, B736698, B736893, B739090, B739091, B795110, B796586, B800827, B801770; PED3-021-350-20/00847536030725, 00847536035218, 00847536036543/B319436, B336571, B379993, B383427, B389708, B400103, B400599, B403612, B407062, B415211, B446722, B483987, B490972, B492403, B494049, B552487, B559741, B591782, B591864, B592074, B622656, B626023, B648708, B684905, B715692, B719565, B719567, B719659, B726923, B728926, B735667, B756330, B765053, B768231, B770237, B770248, B792630; PED3-021-250-20/00847536030442, 00847536034938, 00847536036260/B321465, B337378, B360749, B371070, B412868, B467843, B477487, B488713, B548244, B557602, B644034, B673388, B679822, B684208, B686727, B695486, B698571, B703035, B708080, B709204, B711078, B746067, B757056, B757060, B7610

Distribution

Worldwide distribution: US (nationwide): PR, OK, CA, MD, FL, NY, WA, MI, AL, NV, MN, NJ, IL, OH, AK, CT, TN, MO, TX, ID, AZ, WI, PA, OR, SD, VA, CO, GA, KY, IN, AR, MA, LA, NE, MT, ME, SC, DC, NC, ND, NM, IA, UT, KS, WV, NH, HI, RI. And OUS (foreign) countries of: Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Republic Of Moldova, Montenegro, Morocco, Netherlands, New Zealand, North Macedonia, Northern Ireland, Norway, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam

States: Alaska Alabama Arkansas Arizona California Colorado Connecticut District of Columbia Florida Georgia Hawaii Iowa Idaho Illinois Indiana Kansas Kentucky Louisiana Massachusetts Maryland Maine Michigan Minnesota Missouri Montana North Carolina North Dakota Nebraska New Hampshire New Jersey New Mexico Nevada New York Ohio Oklahoma Oregon Pennsylvania Puerto Rico Rhode Island South Carolina South Dakota Tennessee Texas Utah Virginia Washington Wisconsin West Virginia

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-1246-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.