FDA Medical devices recall ·
LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo¿ RD rainbow Adt 8¿ SpCO… recall
- Recall number
- Z-1184-2025
- Date
- (initiated January 21, 2025)
- Company / brand
- Physio-Control, Inc.
11811 Willows Rd Ne, Redmond, WA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 622
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo¿ RD rainbow Adt 8¿ SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo¿ LNCS-II rainbow¿ DCI 8¿ SpCO, Adult Reusable Sensor REF 11996-000519 Masimo¿ LNCS-II rainbow¿ DCI 8¿ SpCO, Pediatric Reusable Sensor REF 11996-000520 The LIFEPAK¿ 15 Monitor/Defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. RD rainbow" 8¿ SpCO sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO¿), and methemoglobin saturation (SpMet¿) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
Why was it recalled?
Due to an error message that prevents users from utilizing carboxyhemoglobin saturation and methemoglobin saturation sensors on their monitor/defibrillator system.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Catalog Number: 11996-000515 UDI-DI/GTIN code: 00843997010863 Lot Number: 23G79 Catalog Number: 11996-000519 UDI-DI/GTIN code: 00843997015608 Lot Numbers: 23HER 23HPT 23JBN 23JTA Catalog Number: 11996-000520 UDI-DI/GTIN code: 00843997015615 Lot Numbers: 23HNV 23JBP
Distribution
U.S. Nationwide distribution including in the states of CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, MA, MD, MI, MO, MS, NC, NH, NY, OH, PA, SC, TN, TX, VA, WA, WI, and WV.
States: California Colorado Connecticut Florida Georgia Iowa Idaho Illinois Indiana Kentucky Massachusetts Maryland Michigan Missouri Mississippi North Carolina New Hampshire New York Ohio Pennsylvania South Carolina Tennessee Texas Virginia Washington Wisconsin West Virginia
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-1184-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.