FDA Medical devices recall ·
LIFEPAK 15 recall
- Recall number
- Z-2065-2026
- Date
- (initiated April 13, 2026)
- Company / brand
- Physio-Control, Inc.
11811 Willows Rd Ne, Redmond, WA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 206
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001956, 99577-001588, 99577-001958, 99577-001950, 99577-001962, 99577-002178, 99577-001372, 99577-001938, 99577-000045; LIFEPAK 1000, REF: 99425-000025, 99425-000023, 99425-000096; LIFEPAK 35, REF: 99335-000013, 99335-000031; LIFEPAK 20e, REF: 99507-000122
Why was it recalled?
Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Device/UDI/Serial Number: LIFEPAK 15/0088387391164815011543166573, 0088387387169015033143394891, 0088387387169015050743524360, 0088387387169015092843914574, 0088387387204815093043915122, 0088387387169016010444187447, 0088387387162116012044252611, 0088387387169016012644262721, 0088387387169016012644277082, 0088387387169016012744282280, 0088387387171316060244574340, 0088387387171316080444896329, 0088387387169016083144962804, 0088387387170617013145447229, 0088387387169017041245743786, 0088387387171317050945834209, 0088387387171317062845998142, 0088387387171317082846103043, 0088387387169017101046303288, 0088387387171317103146303092, 0088387387162117122146587130, 0088387387171317122646602003, 0088387387171317122846608783, 0088387387171317122946614788, 0088387387171318022246813507, 0088387387169018032946966267, 0088387387171318061947337396, 0088387387169018070547401928, 0088387387171318092747611441, 0088387387169018112947784931, 0088387387169018112947786317, 0088387387171318121847847661, 0088387387171318121847848950, 0088387387708118121947845317, 0088387387708118121947855371, 0088387387169018122047857172, 0088387387169018122047858660, 0088387387171318122847872868, 0088387387171319031248018370, 0088387387169019061248247559, 0088387387168319062748298704, 0088387387169019100848453060, 0088387387171319101448455890, 0088387387171319121848586351, 0088387387171319121948586599, 0088387387168320012848651507, 0088387387171320030548701689, 0088387387708120042848784975, 00883873871713200716488
Distribution
US Nationwide distribution in the states of MI, WA, KY, IN, FL, NC, TX, LA, VA, NY, MA, CO, PA, OH, SC, OK, MN, WV, NJ, WI.
States: Colorado Florida Indiana Kentucky Louisiana Massachusetts Michigan Minnesota North Carolina New Jersey New York Ohio Oklahoma Pennsylvania South Carolina Texas Virginia Washington Wisconsin West Virginia
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2065-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.