FDA Medical devices recall ·
Intact Vascular Tack Endovascular System, to treat vascular dissections… recall
This is an FDA Class I recall — the most serious type.
- Recall number
- Z-1070-2025
- Date
- (initiated January 10, 2025)
- Company / brand
- Spectranetics Corporation
9965 Federal Dr, Colorado Springs, CO, United States - Category
- Medical devices › Class I
- Classification
- Class I
- Status
- Ongoing
- Units affected
- 2,939
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Intact Vascular Tack Endovascular System, to treat vascular dissections with Tack implant(s) following angioplasty: (4F,1.5-4.5mm), 150cm, REF: 154150041; Tack, 4F Gen 1.5, 150cm CE, REF: 154150042; Tack, 4F Gen 1.5, 90cm CE, REF: 154090042 (6F, 3.5 - 6.0mm), 135cm, REF: 156135061; (6F, 4.0 - 8.0mm), 135cm, REF: 206135061; Tack, 6F Gen 2.0, 135cm CE, REF: 206135062; Tack, 6F Gen 1.5, 135cm CE, REF: 156135062; Tack, 6F Gen 2.0, 80cm CE, REF: 206080062; Tack, 6F Gen 1.5, 80cm CE, REF: 156080062
Why was it recalled?
Use of Tack Endovascular system, designed to treat acute dissections of inner wall or lining of an artery by tacking damaged tissue to inner luminal surface, could result in problems with deployment and stability of device in vessel after deployment, which may cause failure to resolve dissection, migration of implant, bailout stenting, reintervention, unintended removal of tack devices, and ischemia.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI-DI: REF/UDI-DI/Batch: 154150041/00863328000189/332905 - 349049, 154150042/328641, REF: 154090042/322101, 156135061/00863328000103/332824 - 336121, 206135061/00850003494043/331281 - 336152, 206135062/328643 - 345567, 156135062/271330 - 283066, 206080062/321083 - 349035, 156080062/302251 - 349034
Distribution
Distribution US nationwide and countries of: NL, DE, NL, AU, IT, CH, GB, and AT
States: Delaware
What does Class I mean?
Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.
Source: FDA (medical devices), recall Z-1070-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.