FDA Medical devices recall ·
VANTAGE TITAN 3T Model MRT-3010/MEXL-3010 recall
- Recall number
- Z-1018-2026
- Date
- (initiated December 19, 2025)
- Company / brand
- Canon Medical System, USA, INC.
2441 Michelle Dr, Tustin, CA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 22 systems
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
VANTAGE TITAN 3T Model MRT-3010/MEXL-3010
Why was it recalled?
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Model Number: MRT-3010/MEXL-3010 UDI-DI code: 04987670101649 CMSC Model Number/Model Number/Serial Numbers/SID/UDI code: MRT-3010/A5 MEXL-3010/A5 A5D14X2031 30000041 MRT-3010/A5 MEXL-3010/A5 A5D13Y2025 30001708 MRT-3010/A5 MEXL-3010/A5 A5D13Y2026 30005894 MRT-3010/A5 MEXL-3010/A5 A5G1692039 30007620 (01)04987670101649(21)A5G1692039 MRT-3010/A5 MEXL-3010/A5.007 A5B11Y2006 30019175 MRT-3010/A7 MEXL-3010/S7 S7A1642006 30024266 (01)04987670101649(21)A7A1642001 MRT-3010/A5 MEXL-3010/A5 A5G1692038 30027318 (01)04987670101649(21)A5G1692038 MRT-3010/A5 MEXL-3010/A5 A5G1642036 30029482 (01)04987670101649(21)A5G1642036 MRT-3010/A5 MEXL-3010/A5.007 A5F1562034 30050314 MRT-3010/A5 MEXL-3010/A5 A5B11Z2007 303354 MRT-3010/A5 MEXL-3010/A5 A5C1252012 336719 MRT-3010/A5 MEXL-3010/A5 A5D1312019 350553 MRT-3010/A5 MEXL-3010/A5 A5B1192003 350579 MRT-3010/A5 MEXL-3010/A5 A5C12X2016 350918 MRT-3010/A5 MEXL-3010/A5 A5C12Y2017 351122 MRT-3010/A5 MEXL-3010/A5 A5C1212009 351387 MRT-3010/A5 MEXL-3010/A5 A5B11Y2005 351452 MRT-3010/A5 MEXL-3010/A5 A5C1262013 356865 MRT-3010/A5 MEXL-3010/A5 A5C1292015 378015 MRT-3010/A5 MEXL-3010/A5 A5C1412028 383946 MRT-3010/A5 MEXL-3010/A5 A5E1522033 389630 MRT-3010/A5 MEXL-3010/A5 A5D1362021 390397
Distribution
U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI.
States: California Colorado Florida Iowa Illinois Indiana Kansas Maryland Missouri Montana Nevada New York Ohio Oregon Pennsylvania Puerto Rico Texas Virginia Wisconsin
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-1018-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.