FDA Medical devices recall ·

ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM recall

Recall number
Z-0289-2027
Date
(initiated October 12, 2026)
Company / brand
Canon Medical System, USA, INC.
2441 Michelle Dr, Tustin, CA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
15 SYSTEMS
Hazard tags
Software
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000C

Why was it recalled?

Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

DI-04987670100192

Distribution

US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, HI, IL, LA, MA, MI, MN, MS, NC, NV, OH, OR, TX.

States: Alabama Arkansas Arizona California Florida Georgia Hawaii Illinois Louisiana Massachusetts Michigan Minnesota Mississippi North Carolina Nevada Ohio Oregon Texas

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0289-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.