FDA Medical devices recall ·

BD Alaris infusion Pump Module 8100 recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-0430-2026
Date
(initiated October 17, 2025)
Company / brand
CareFusion 303, Inc.
10020 Pacific Mesa Blvd, San Diego, CA, United States
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
2,574,396
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100, REF: 49000270. Used with the following new and updated documents: BD Alaris Infusion System Inspection Safety Reminder, Part: BD-163124, Revision: 00; Cleaning and Disinfecting Procedure, Part: P00000864; Cleaning and Disinfecting Quick Reference Guide

Why was it recalled?

If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

All Alaris Pump Modules distributed after June 19, 2004. REF/UDI-DI: 8100/10885403810015, 10885403810046, 10885403810039, 10885403222054, 10885403517723, 10885403213212, 10885403211423; 49000437/10885403471018; 49000270/10885403247743

Distribution

Worldwide distribution. US nationwide including Puerto Rico, Guam, American Samoa, Northern Mariana Islands; and countries of SG, BH, AE, JP, CA, IT, QA, PH, DE, TW, PK, KR, BW, SA, MY, IL, NZ, AU, IN, GB, CN, NL, FR, ZA, BE, MX, CH, GI, KW, and TR.

States: California Delaware Illinois Indiana Puerto Rico

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-0430-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.