FDA Medical devices recall ·

CellMek SPS Sample Preparation System recall

Recall number
Z-0412-2026
Date
(initiated September 8, 2025)
Company / brand
Beckman Coulter, Inc.
11800 Sw 147th Ave, Miami, FL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
108 units
Hazard tags
Software
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

CellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU) revisions AD and prior. These CellMek devices include Cassette E.

Why was it recalled?

Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

All serial numbers between BE43003 through BJ18014, manufactured from 12/03/2021 to 04/15/2025. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU) revisions AD and prior. These CellMek devices include Cassette E.

Distribution

Worldwide - US Nationwide distribution in the states of AL, AR, FL, IL, IN, MA, ME, MI, MN, MO, NY, OR, TN, UT, WA and the countries Australia, Austria, Belgium, Brazil, Canada, China, France, Germany, Greece, Hong Kong, Israel, Italy, Japan, Korea, Republic of, Mexico, Morocco, Netherlands, Poland, Portugal, Puerto Rico, South Africa, Spain, Switzerland, United Kingdom of Great Britain and Northern Ireland.

States: Alabama Arkansas Florida Illinois Indiana Massachusetts Maine Michigan Minnesota Missouri New York Oregon Puerto Rico Tennessee Utah Washington

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0412-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.