FDA Medical devices recall ·
OCS Liver Perfusion Set Model recall
- Recall number
- Z-0381-2027
- Date
- (initiated May 15, 2026)
- Company / brand
- TransMedics, Inc.
200 Minuteman Rd Ste 302, Andover, MA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 18
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
OCS Liver Perfusion Set Model/Catalog Number: 3200 Software Version: rev 3 OCS Liver Perfusion Module included in the Liver Perfusion Set used for organ transport.
Why was it recalled?
The Perfusion Module may have an unprogrammed Front End Board (FEB) which will not allow the system to proceed beyond the initial startup process. The OCS system will display an error message saying Unsupported Perfusion Module Present and will prevent further system use.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
1. Perfusion Set SN: 4470-4-24374; Description: OCS Liver Perfusion Set; EXP Date: 11/10/27; UDI: *+B00944705/$$+74470-4-24374/14D202711100*; 2. Perfusion Set SN: 4470-4-23845; Description: OCS Liver Perfusion Set; EXP Date: 10/23/27; UDI: *+B00944705/$$+74470-4-23845/14D202710235*; 3. Perfusion Set SN: 4470-4-24226; Description: OCS Liver Perfusion Set; EXP Date: 11/10/27; UDI: *+B00944705/$$+74470-4-24226/14D20271110$*; 4. Perfusion Set SN: 4470-4-24559; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: *+B00944705/$$+74470-4-24559/14D202711026*; 5. Perfusion Set SN: 4470-4-24636; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: *+B00944705/$$+74470-4-24636/14D202711022*; 6. Perfusion Set SN: 4470-4-24508; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: *+B00944705/$$+74470-4-24508/14D202711020*; 7. Perfusion Set SN: 4470-4-24384; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: *+B00944705/$$+74470-4-24384/14D202711022*; 8. Perfusion Set SN: 4470-4-23182; Description: OCS Liver Perfusion Set; EXP Date: 10/23/27; UDI: *+B00944705/$$+74470-4-23182/14D20271023%*; 9. Perfusion Set SN: 4470-4-23650; Description: OCS Liver Perfusion Set; EXP Date: 10/19/27; UDI: *+B00944705/$$+74470-4-23650/14D202710194*; 10. Perfusion Set SN: 4470-4-23593; Description: OCS Liver Perfusion Set; EXP Date: 10/19/27; UDI: *+B00944705/$$+74470-4-23593/14D20271019A*; 11. Perfusion Set SN: 4470-4-23471; Description: OCS Liver Perfusion Set; EXP Date:
Distribution
Worldwide - US Nationwide distribution in the states of CA, FL, IL, MA, MN, NC, NE, NY, TX and the country of Serbia;.
States: California Florida Illinois Massachusetts Minnesota North Carolina Nebraska New York Texas
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0381-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.