FDA Medical devices recall ·

OCS Liver System for organ transport OCS Liver Perfusion Set Model recall

Recall number
Z-0389-2027
Date
(initiated November 20, 2025)
Company / brand
TransMedics, Inc.
200 Minuteman Rd Ste 302, Andover, MA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1179 (Liver Perfusion Sets)
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

OCS Liver System for organ transport OCS Liver Perfusion Set Model/Catalog Number: 3200 Product Description: The recalled product is a standard off the shelf syringe. It is used as part of solution delivery cassette (SDS Cassette) to be used with the perfusion module of OCS Liver System. OCS Liver Perfusion Sets within the serial number range between 4470-3-18076 and 4470-3-20343 may contain SDS cassettes within the impacted lot range of 4459-6-737116 to 4459-6-739965. Each Liver Perfusion Set contains three SDS cassettes. The BD syringe is a component of SDS Cassette PN 100004459, which is an accessory of the Liver Perfusion Set.

Why was it recalled?

Faulty Solution Delivery System (SDS) Cassettes within the device set were assembled with syringes that were manufactured with an atypical lubricant, which results in sub-par performance and has the potential to cause leaks.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

OCS Liver Perfusion Set Serial Numbers from 4470-3-18076 to 4470-3-20343;

Distribution

Worldwide - US Nationwide distribution in the state of NE and the country of Italy.

States: Nebraska

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0389-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.