FDA Medical devices recall ·
RF HEAD 24967 PATIENT CONNECTOR and RF HEAD 24967 EUR MKT… recall
- Recall number
- Z-0363-2027
- Date
- (initiated September 24, 2026)
- Company / brand
- Medtronic, Inc.
8200 Coral Sea St Ne, Mounds View, MN, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 11741 untis
- Hazard tags
- Software
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
RF HEAD 24967 PATIENT CONNECTOR and RF HEAD 24967 EUR MKT, communicates with Medtronic cardiac devices, Product Number 24967
Why was it recalled?
Bluetooth disconnections may occur between the device manager software and its associated accessories. Bluetooth disconnections during a SmartSync patient session may lead to harm due to undesirable or incomplete parameter programming.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI-DI: 00763000060435, 00763000060428, 00763000060442, 00763000060459, 00763000288242, 00763000257583, 00763000257613, 00763000317799, 00763000257576, 00763000317775, 00763000208394, 00763000424374, 00763000688479, 00763000688509, 00763000688516, 00763000688530, 00763000688554, 00763000688493, 00763000688547, 00763000688486; Serial Numbers: RFA001598A RFA001602A RFA001786A RFA001826A RFA002016A RFA002095A RFA002591A RFA003113A RFA003146A RFA003306A RFA005031A RFA005097A RFA005103A RFA008830A RFA014171A RFA014218A RFA014252A RFA014258A RFA014278A RFA014286A RFA014342A RFA014609A RFA014653A RFA014885A RFA014952A RFA014996A RFA015015A RFA016648A RFA016815A RFA016817A RFA017064A RFA017468A RFA017880A RFA017906A RFA017907A RFA017929A RFA017947A RFA018801A RFA019059A RFA021437A RFA021450A RFA021477A RFA021618A RFA021626A RFA021652A RFA021661A RFA036107A RFA036123A RFA036190A RFA036411A RFA036460A RFA037606A RFA037647A RFA037648A RFA037650A RFA037794A RFA038977A RFA039112A RFA039113A RFA040362A RFA040386A RFA040584A RFA040632A RFA040655A RFA040688A RFA040689A RFA040765A RFA040802A RFA040823A RFA040848A RFA040850A RFA040853A RFA040905A RFA040911A RFA040918A RFA041045A RFA041053A RFA041211A RFA041213A RFA041375A RFA041376A RFA041383A RFA041384A RFA041450A RFA041455A RFA041457A RFA041483A RFA041511A RFA041514A RFA041515A RFA041518A RFA041537A RFA041559A RFA041568A RFA041572A RFA041573A RFA041600A RFA042036A RFA042051A RFA042055A RFA042056A RFA042124A RFA042131A RFA042132A RFA042141A R
Distribution
Worldwide - US Nationwide distribution.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0363-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.