FDA Medical devices recall ·

CareLink SmartSync Device Manager App, a tablet-based platform used to… recall

Recall number
Z-0361-2027
Date
(initiated September 24, 2026)
Company / brand
Medtronic, Inc.
8200 Coral Sea St Ne, Mounds View, MN, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
38175 systems
Hazard tags
Software
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

CareLink SmartSync Device Manager App, a tablet-based platform used to program and interrogate compatible Medtronic CIEDs, Product Number M01A02

Why was it recalled?

Bluetooth disconnections may occur between the device manager software and its associated accessories. Bluetooth disconnections during a SmartSync patient session may lead to harm due to undesirable or incomplete parameter programming.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI: NA; Serial Numbers: C2TM4QV0XC DL269KH953 DVP79R6KV6 FDG69FQX19 FHHG9CHX3H FVFVMVPK7N FVQYCR34T1 FYJPFH6099 G19LC9X7J7 G2VQP97D2J GR4R69G4VN GW7TFH7HQ0 GX44D2GDP4 H3G0Q6W9T7 H3QGWTYW42 H9GX6NHL45 HD2104VHK2 HD972G7JFV HK212M19PD HPH6FQFHF9 J4G16FXW05 JGYXLTCXLW K6HTFXLW2T KD0MV7CVQG KQL96W3WGF MGF4DJCVFL Q76YQPKJQK QXGQV69765 RH1XG26WGV V669K664R6 X2XDNVV5CF C2TR7WKTWW C44L16FCW7 C52X07MXR7 C5Y6H49KH6 C69TG9WDJV C69W9N6PWF C6VT6QX05N C9W7NX7X11 CFQT0F4N2T CH02T6FTFJ CJ2VX4PCQT CJF6VNT3HT CJR4Y6920X CM44FWL9G7 CP63R629G3 CQ94K67LJF CW3KFX02JG CW7XHL6MPY CW9LQ9593X CWJX3439G5 CWNW7QK7YW CWR0W7J4VL CWRR649J63 CWVJ60WDCC CWVXWHWG17 CWYMJQKVDN CX99YN7PL2 CXGCY70L60 CXWJLXQYX6 D04T9VT46M D0JFR44TD2 D4070M21J1 D494VRK60H D617Y162PJ D6VF5YXXH1 D70KX7QJ22 D723GV1WQW D7K79L70F6 D7LG73X9JW D7RJ2T9KC9 D96NK032FY DH94DT62V4 DTD172G4YP DY63799N29 F062YWR7V3 F2NL42C2MM F3WV20DVTJ F3Y0R2YGQT F4KNXN3609 F4XWWJ5TXM F5VX9GHG0W F67W79C3FH F67XVY2M27 F6C6667WJ6 F6P9MNVDYR F7NP73CN7H F7NYPL2MKC FD90MH6YVY FH497W575J FHN2F9F614 FHYYG7FDJP FK6JX9D61J FKPFXKGWC4 FLYF7YXF69 FMHXCDJJMG FNXQ10D4RW FPQT4VT2JN FT6HQK60D7 FVJG2N26J1 FWRJF0VMR9 FWVYMJH9TR FX1Y74DLC4 FX2337J9GQ FX76D726VH FY4016742W FY90KH799T FYD90QM3W0 G07VQ2G2XK G0HK4Q75WV G170JGHLQL G1C96DL6C9 G1CF04Q3H9 G2DT932HNY G2PX4QHHG3 G46K22XXN9 G4TYJ0QM7X G4WCV2RYR7 G5X4CXQYR3 G6J29221WX G9C979CX95 GH4V46FGPG GJ2KD063F7 GJ9YWDKQCV GJQMT4K3DT GL91WL5X7M GMK3JWNR94 GMKPJ67Y4P GN7N7K7XYM GN9216D3R4 GNPRYX0XWP GP24M66J3Y GPR6RPFCJ7 GQ0CV4CF0

Distribution

Worldwide - US Nationwide distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0361-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.