FDA Medical devices recall ·

VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7 recall

Recall number
Z-0327-2026
Date
(initiated October 16, 2025)
Company / brand
Canon Medical System, USA, INC.
2441 Michelle Dr, Tustin, CA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
2 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.

Why was it recalled?

Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7. Serial Numbers: B7A2462002, B7A2522014.

Distribution

US distribution to Ohio and Nevada

States: Nevada Ohio

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.