FDA Medical devices recall ·

LOGIQ P10 series with software version R4.5.7 Model Number 5877534 recall

Recall number
Z-0155-2026
Date
(initiated September 18, 2025)
Company / brand
GE Medical Systems, LLC
3000 N Grandview Blvd, Waukesha, WI, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
71 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

LOGIQ P10 series with software version R4.5.7 Model Number 5877534

Why was it recalled?

The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Model 5877534 UDI-DI: 00195278464644, Serial Numbers: LPX441358 LPX441335 LPX440894 LPX440598 LPX440898 LPX441357 LPX441328 LPX441402 LPX441407 LPX441405 LPX441326 LPX441126 LPX441406 LPX441403 LPX441404 LPX441439 LPX441440 LPX441327 LPX441043 LPX441399 LPX441370 LPX441422 LPX441369 LPX441001 LPX441447 LPX440990 LPX441374 LPX441376 LPX441428 LPX441453 LPX441427 LPX440999 LPX441373 LPX441420 LPX441384 LPX441375 LPX441421 LPX441424 LPX441371 LPX441425 LPX441426 LPX441367 LPX441368 LPX441377 LPX441382 LPX441455 LPX441423 LPX441430 LPX441018 LPX441417 LPX441437 LPX441445 LPX441265 LPX441446 LPX441412 LPX441451 LPX441462 LPX441448 LPX441450 LPX441475 LPX441457 LPX441444 LPX490014 LPX441099 LPX441452 LPX441441 LPX441442 LPX441408 LPX441343 LPX441517 LPX441501

Distribution

US Nationwide. Global Distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0155-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.