FDA Medical devices recall ·

Boston Scientific AVVIGO+ Multi-Modality Guidance System… recall

Recall number
Z-0107-2025
Date
(initiated September 5, 2024)
Company / brand
Boston Scientific Corporation
1 Scimed Pl, Maple Grove, MN, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
8 units
Hazard tags
Software
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM EVAL US; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which includes intravascular ultrasound (IVUS). When used with a peripheral vascular imaging catheter, the system utilizes the IVUS modality for ultrasound examination of intravascular pathology.

Why was it recalled?

Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Material Number H7492493120D0, GTIN/UDI/DI 00191506033149, Batch/Serial Numbers: 102899658, 103225132, 103246673, 103647916, 104043675, 104072168, 104258818, 104512059

Distribution

US Nationwide distribution in Puerto Rico.

States: Puerto Rico

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0107-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.