FDA Medical devices recall ·
CHARLOTTE CLAW Screw recall
- Recall number
- Z-3334-2026
- Date
- (initiated August 18, 2026)
- Company / brand
- Wright Medical Technology, Inc.
11500 memphis arlington rd, arlington, TN, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 34
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System
Why was it recalled?
Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI-DI: 00840420109705, Lot: 1826666, Distributed: 18-Aug-2025 through 09-Apr-2026
Distribution
US Nationwide distribution in the states of FL, TX, PA, CA, ID, MI, OK, TN, NJ, WI, CO.
States: California Colorado Florida Idaho Michigan New Jersey Oklahoma Pennsylvania Tennessee Texas Wisconsin
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-3334-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.