FDA Medical devices recall ·
Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger… recall
- Recall number
- Z-3327-2026
- Date
- (initiated July 17, 2026)
- Company / brand
- Spectrum Medical Srl
Via Di Mezzo 23, Mirandola, Italy - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 9448
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 W Model/Catalog Number: HX55V-S0W Software Version: N/A Product Description: The Quantum PureFlow Standard Heat Exchanger is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration with a maximum flow rate of 8 l/min. Quantum PureFlow Standard Heat Exchanger is a single use device which allows heating/cooling the blood during extracorporeal circulation. The device is sold sterile in a blister pack in a multiple box of 4 units. Device surfaces in contact with blood are treated with a phosphorylcholine-based biocompatible coating. The device has been designed to be powered by heaters coolers that use water or a glycol-based heat transfer fluid (HTF). Component: N/A
Why was it recalled?
Due to blood-to-glycol leak in the heat exchanger.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Lot Code: 18051160302205 (HX55V-S0W)
Distribution
US Nationwide distribution in the states of AL, CA, FL, KS MA, MD, MI, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, WI.
States: Alabama California Florida Kansas Massachusetts Maryland Michigan Missouri North Carolina New Jersey New York Ohio Oregon Pennsylvania South Carolina Tennessee Texas Virginia Wisconsin
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-3327-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.