FDA Medical devices recall ·

ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part… recall

Recall number
Z-3034-2026
Date
(initiated August 6, 2026)
Company / brand
IMRIS Imaging Inc
1230 Chaska Creek Way Ste 100, Chaska, MN, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
13 units
Hazard tags
Injury risk
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600

Why was it recalled?

The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

114093-000 - UDI-DI: 00857534006356, Serial Numbers: 10005695, 10002998, 10002997, 10005456, 10005612, 10004372, 10005613, 10002841, 10011119; 114093-600 - UDI-DI: 00857534006462, Serial Numbers: 10002996, 10004515, 10007411, 10007376

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-3034-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.