FDA Medical devices recall ·

U2 Total Knee System Product Name: PSA Box Trial, L… recall

Recall number
Z-2883-2026
Date
(initiated May 20, 2026)
Company / brand
United Orthopedic Corp.
No. 16 Luke 1st Rd., Kaohsiung, Taiwan
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
2
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: U2 Total Knee System Product Name: PSA Box Trial, L, #1-6 and PSA Box Trial, R, #1-6 Model/Catalog Number: 9323-5511 thru 9323-5516 and 9323-5521 thru 9323-5526 Software Version: N/A Product Description: Knee femur prosthesis trial, reusable Component: No, N/A

Why was it recalled?

Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Product Number; UDI-DI Code; Lot Code: Product Number 9323-5511; UDI-DI 04719886906929; Lot number 25FB54 Product Number 9323-5512; UDI-DI 04719886906936; Lot number 25FB55 Product Number 9323-5513; UDI-DI 04719886906943; Lot number 25FB56 Product Number 9323-5514; UDI-DI 04719886906950; Lot number 25FB57 Product Number 9323-5515; UDI-DI 04719886906967; Lot number 25FB58 Product Number 9323-5516; UDI-DI 04719886906424; Lot number 25FB59 Product Number 9323-5521; UDI-DI 04719886906974; Lot number 25FB29 Product Number 9323-5522; UDI-DI 04719886906981; Lot number 25FB30 Product Number 9323-5523; UDI-DI 04719886906998; Lot number 25FB31 Product Number 9323-5524; UDI-DI 04719886907001; Lot number 25FB32 Product Number 9323-5525; UDI-DI 04719886907018; Lot number 25FB33 Product Number 9323-5526; UDI-DI 04719886907025; Lot number 25FB34

Distribution

US Nationwide distribution in the states of Texas and Michigan.

States: Michigan Texas

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2883-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.